It is designed to emit a precisely defined hydrogen leak rate so that helium sniffer leak detectors can be verified and calibrated. The device is a pressurized, precision-engineered calibrated leak that emits 100% hydrogen at a known, extremely small leak rate:
Helium sniffer leak detectors must be periodically checked against a known, certified leak to ensure:
The calibrated leak provides a traceable reference point, so the detector can confirm it is reading the correct leak rate.
Technicians attach or present the calibrated leak to the sniffer probe of the ASM 306 S. The detector should read the known leak rate; if not, the instrument may need adjustment or service.
The Hydrogen Leak is the “test leak” that ensures your helium leak detector is working correctly, not the detector itself. (For Installation Instuctions and Operation see .PDF Below in Downloads.)